FDA Recall Terminated

Kenex Radiation Shield installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging systems.

Recall: Z-1441-2015 · Initiated March 9, 2015

Recall

Recall Number
Z-1441-2015
Event Number
70749
Firm
GE Healthcare
FEI Number
2126677
Product Code
KPY
Status
Terminated
Root Cause
Equipment maintenance
Initiated
March 9, 2015
Posted
April 13, 2015
Terminated
December 21, 2015
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Kenex Radiation Shield installed with GE Healthcare Interventional, Surgery, CT and X-Ray Imaging systems.

Reason

GE Healthcare has recently become aware of an improper installation of a ceiling suspended pendent supporting a radiation shield and/or surgical lamp. The pendent system was not installed per specified installation requirements. There has been a reported incident of a radiation shield falling from the ceiling causing a minor injury.

Action

GE Healthcare sent an Urgent Medical Device Correction letter dated March 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in USA contact 1-800-437-1171. For other countries contact your local GE Healthcare Service Representative.

Distribution

Worldwide Distribution - US Distribution to the states of : CA and MA., and to the countries of : SPAIN, INDIA, JAPAN, AZERBAIJAN, BULGARIA, BELARUS, SWITZERLAND, CZECH REPUBLIC, GERMANY, EGYPT, FINLAND, FRANCE, UNITED KINGDOM, GUYANA, IRELAND, ISRAEL, ITALY, NIGERIA and NORWAY.

Quantity

797 (3 US, 794 OUS).