FDA Recall
Terminated
Anestar Anesthesia Delivery System
Recall: Z-1441-04
·
Initiated June 24, 2003
Recall
- Recall Number
- Z-1441-04
- Event Number
- 29545
- Firm
- Datascope Corp
- FEI Number
- 2221819
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 24, 2003
- Posted
- September 9, 2004
- Terminated
- October 3, 2006
- Address
- 800 MacArthur Blvd, Mahwah, NJ, 07430-0619
Description
Anestar Anesthesia Delivery System
Reason
Failure of the Anestar AC Main Power Switch. Battery should last 30 minutes, then ventilation and monitoring will fail.
Action
A Field Correction action was initiated in June, 2003 and completed by the service reps in April, 2004.
Distribution
The units were distributed to 29 hospitals nationwide. There are no US Government military or civilian customers.