FDA Recall Terminated

Aqueduct Hematology Workflow Manager, software

Recall: Z-1440-05 · Initiated August 10, 2005

Recall

Recall Number
Z-1440-05
Event Number
32931
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
JQP
Status
Terminated
Root Cause
Other
Initiated
August 10, 2005
Terminated
April 3, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Aqueduct Hematology Workflow Manager, software

Reason

Potentail for misidentification of historical patient results within the Aqueduct Hematology Workflow Manager software. In instances where the workflow manager software receives the same patient identifier for more than one patient the Aqueduct Workflow Manager software merges all results into one Patient File. The may cause test results to be associated with the wrong patient.

Action

A Product corrective action letter to be mailed the week of August 8, 2005, to all customers directing that each patient is identified with a truly unique patient identifier.

Distribution

AK, AZ, CA, DE, FL, IN, KS, MA, MI, MN, MO, NC, NY, OH, PA, TX, VA, WA

Quantity

36