FDA Recall Terminated

NobelReplace CC PMC RP 5.0x10mm, Article No. 37296.

Recall: Z-1439-2015 · Initiated March 9, 2015

Recall

Recall Number
Z-1439-2015
Event Number
70821
Firm
Nobel Biocare Usa Llc
FEI Number
1000125598
Product Code
DZE
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 9, 2015
Posted
April 13, 2015
Terminated
December 30, 2015
Address
22715/22725 Savi Ranch Pkwy, Yorba Linda, CA, 92887

Description

NobelReplace CC PMC RP 5.0x10mm, Article No. 37296.

Reason

A material used in the packaging of the dental implants may impart a sticky substance on the top surfaces of the implants and in some cases on the first threads.

Action

Nobel Biocare sent an Important - Field Correction letter dated March 6, 2015, to all affected customers and another recall letter dated dated March 10, 2015, to all affected customers to inform them that Nobel Biocare Quality Management Systems observed an issue with a material used in the packaging of their implants. The firm's internal analysis has confirmed that this material is not according to their standards. The letters provide the problems identified and the actions to be taken. Customers were instructed to return the affected product(s) not yet used to Nobel Biocare to receive a replacement product. Customers were also instructed to complete the attached Acknowledgement form and fax it to 714-282-5073 or email it to [email protected]. Customers with questions were instructed to contact their customer service department. For questions regarding this recall call 714-282-4800.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Austria, Australia, Belgium, Bulgaria, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Lithuania, Luxembourg, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Sweden, Turkey, and United Arab Emirates.

Quantity

53 units