FDA Recall Terminated

Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging systems. Component material number 7559375 Upper Body radiation shield, Artis T.

Recall: Z-1439-2009 · Initiated May 1, 2009

Recall

Recall Number
Z-1439-2009
Event Number
51947
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IZI
Status
Terminated
Root Cause
Other
Initiated
May 1, 2009
Posted
June 16, 2009
Terminated
September 1, 2009
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging systems. Component material number 7559375 Upper Body radiation shield, Artis T.

Reason

Improper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.

Action

The recalling firm has decided to inspect all of the installed units for this condition and where necessary replace the missing clip. The recalling firm has issued an Update Instructions AX011/09/S dated 03.09, which provides instructions on how to perform a visual inspection of the Mavig Portegra 2 arm for the presence of the safety clip and the installation instructions on how to properly install the clip if it is missing.

Distribution

Nationwide Distribution.

Quantity

427 units