Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging systems. Component material number 7559375 Upper Body radiation shield, Artis T.
Recall
- Recall Number
- Z-1439-2009
- Event Number
- 51947
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 1, 2009
- Posted
- June 16, 2009
- Terminated
- September 1, 2009
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Mavig Portegra 2 which is an accessory to the Siemens angiographic, fluoroscopic and radiographic imaging systems. Component material number 7559375 Upper Body radiation shield, Artis T.
Improper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
The recalling firm has decided to inspect all of the installed units for this condition and where necessary replace the missing clip. The recalling firm has issued an Update Instructions AX011/09/S dated 03.09, which provides instructions on how to perform a visual inspection of the Mavig Portegra 2 arm for the presence of the safety clip and the installation instructions on how to properly install the clip if it is missing.
Nationwide Distribution.
427 units