CELL-DYN Ruby System: List Number: 08H6701; Hematology: The CELL-DYN Ruby System is a multi-parameter automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories
Recall
- Recall Number
- Z-1438-2017
- Event Number
- 76481
- Firm
- Abbott Laboratories
- FEI Number
- 2919069
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- February 13, 2017
- Posted
- March 9, 2017
- Terminated
- March 6, 2018
- Address
- 4551 Great America Pkwy, Santa Clara, CA, 95054-1208
Description
CELL-DYN Ruby System: List Number: 08H6701; Hematology: The CELL-DYN Ruby System is a multi-parameter automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories
The Pump Relay Printed Circuit Board Assembly (PCBA) used in the CELL-DYN Ruby instrument may prematurely fail and lead to instrument stoppage
Product correction letters were sent to all affected customers on February 13, 2017 by traceable method. Letters provided information on the reason for the correction with all affected serial numbers listed. Abbott will be contacting each location to schedule replacement of the PCBA Pump Relay Board. Update: five additional serial numbers were added. Correction letters were sent to new consignees - 3 in Brazil and 2 in Mexico.
US and Afghanistan, Brazil, Egypt, Tanzania, Hungary, Iraq, Kenya, Netherlands, Russia, Romania Vietnam, , United Kingdom, Saudi Arabia, Philippines, Lebanon, Mexico, Japan, Jordan, India, Indonesia, Turkey, UAE, Gaza & Jericho, Germany, Bulgaria, Chile, Argentina, Luxembourg.
116