FDA Recall Terminated

REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head Metal Lock Pin is to be inserted into the anterior hole of the tibial bearing yoke to secure the OSS Axle in the correct position within the OSS Femoral component assembly.

Recall: Z-1438-2012 · Initiated March 20, 2012

Recall

Recall Number
Z-1438-2012
Event Number
61089
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
KRO
Status
Terminated
Root Cause
Component design/selection
Initiated
March 20, 2012
Posted
April 19, 2012
Terminated
February 7, 2013
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

REF. CP113456, MAK OSS No Head Lock Pin, NO HEAD / WITH PLUG CO-CR-MO ALLOY/ARCOM UHMWPE STERILE, R, Biomet Orthopedics Warsaw, IN 46581. The MAK OSS No Head Metal Lock Pin is to be inserted into the anterior hole of the tibial bearing yoke to secure the OSS Axle in the correct position within the OSS Femoral component assembly.

Reason

Biomet initiated this action after receiving complaints of the MAK OSS No Head Lock Pin dislodging from its intended position (post-initial surgery). An investigation was initiated and the available lots of the Lock Pin were evaluated and measured against print specifications.

Action

Biomet sent an "URGENT MEDICAL DEVICE SAFETY ALERT" letter dated March 20, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to continue monitoring implanted patients for disassociation of the pin component and report to Biomet. The Alert included a complete listing of device invoice history and patient information (if supplied). Contact the firm for confirmation of receipt of the notice and for questions at 574-372-3983.

Distribution

Distributed in California and Pennsylvania.

Quantity

64 units distributed