FDA Recall Terminated

Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.

Recall: Z-1436-2012 · Initiated January 19, 2009

Recall

Recall Number
Z-1436-2012
Event Number
52013
Firm
Baxa Corporation
FEI Number
1419106
Product Code
LNX
Status
Terminated
Root Cause
Labeling design
Initiated
January 19, 2009
Posted
May 25, 2012
Terminated
September 12, 2012
Address
14445 Grasslands Dr, Englewood, CO, 80112

Description

Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.

Reason

Ordering salt-based ingredients on an ion-based template resulted in medication errors when calculating amounts of ingredients required.

Action

Baxa Corporation sent a "Safety Alert" letter dated January 19, 2009 via FedEx to all affected customers. The letter described the affected product, problem, and actions to be taken. Customers were instructed to ensure that users create either a salt-based template OR an ion-based template for ordering, and that both methods of ordering are not used in one template. Customers were provided a list of safe practices to follow on the enclosed ISMP Safety Alert. Customers were advised upon receipt of the Safety Alert that Baxa Technical Support will review and address any issues with the customer on their configuration setup for calcium and magnesium. For questions contact Baxa Technical Support at 1-800-678-2292.

Distribution

Worldwide Distribution - US (Nationwide) and the countries of: Canada, China, Australia, Panama and Columbia.

Quantity

667 units