FDA Recall Terminated

Roche/Hitachi AST (ASAT/GOT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1928511 (11928511216).

Recall: Z-1435-04 · Initiated August 20, 2004

Recall

Recall Number
Z-1435-04
Event Number
29893
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
CIT
Status
Terminated
Root Cause
Other
Initiated
August 20, 2004
Posted
September 2, 2004
Terminated
November 5, 2004
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Roche/Hitachi AST (ASAT/GOT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1928511 (11928511216).

Reason

Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.

Action

Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.

Distribution

Nationwide.

Quantity

757non-expired kits.