Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1877470 (11877470216).
Recall: Z-1434-04
·
Initiated August 20, 2004
Recall
- Recall Number
- Z-1434-04
- Event Number
- 29893
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- CKH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 20, 2004
- Posted
- September 2, 2004
- Terminated
- November 5, 2004
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1877470 (11877470216).
Reason
Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.
Action
Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.
Distribution
Nationwide.
Quantity
728 non-expired kits.