FDA Recall Terminated

COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.

Recall: Z-1433-04 · Initiated August 12, 2004

Recall

Recall Number
Z-1433-04
Event Number
29880
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
GKZ
Status
Terminated
Root Cause
Other
Initiated
August 12, 2004
Posted
September 2, 2004
Terminated
April 26, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.

Reason

Beckman Coulter has confirmed an issue associated with the predilute mode of the COULTER LH 500 Hematology Analyzer. The workstation may display, transmit, and / or print an erroneous result or an incorrect dilution factor with a predilute sample result.

Action

Customers were notified via U.S. mail by a Product Corrective Action Letter (PCA) dated August 12, 2004. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occurring.

Distribution

Nationwide & Canada

Quantity

222