Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.
Recall
- Recall Number
- Z-1432-2010
- Event Number
- 55103
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- JAI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 11, 2007
- Posted
- April 23, 2010
- Terminated
- December 28, 2010
- Address
- 4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011
Description
Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.
185 units