FDA Recall Terminated

Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.

Recall: Z-1432-2010 · Initiated January 11, 2007

Recall

Recall Number
Z-1432-2010
Event Number
55103
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
JAI
Status
Terminated
Root Cause
Other
Initiated
January 11, 2007
Posted
April 23, 2010
Terminated
December 28, 2010
Address
4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011

Description

Precise Treatment Table Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use.

Reason

In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.

Action

Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.

Distribution

Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.

Quantity

185 units