FDA Recall Terminated

Mallinckrodt Satin-Slip Intubation Stylet, 6 Fr., single use, Model number 85863,shipped in cases of 20

Recall: Z-1431-06 · Initiated July 12, 2006

Recall

Recall Number
Z-1431-06
Event Number
35879
Firm
Nellcor Puritan Bennett
FEI Number
2936999
Product Code
BSR
Status
Terminated
Root Cause
Other
Initiated
July 12, 2006
Posted
August 24, 2006
Terminated
January 10, 2007
Address
4280 Hacienda Dr, Pleasanton, CA, 94588-2719

Description

Mallinckrodt Satin-Slip Intubation Stylet, 6 Fr., single use, Model number 85863,shipped in cases of 20

Reason

Product's outer sheath material may separate from the wire interior during use and may lodge in the endotracheal tube or enter the patient's airway and interfere with ventilation.

Action

The firm has issued notification letters via USPS certified mail with return receipt, requesting that the recipients who have further distributed the product forward a copy of that letter. The firm intends to contact non-responding companies beginning 30 days after the initial notification is sent.

Distribution

The product has been distributed to 454 consignees throughout the United States, Guam and Japan.

Quantity

30,366 cases of 20 units per case were distributed worldwide