FDA Recall Open, Classified

HAMILTON-C6, REF: 160021

Recall: Z-1429-2023 · Initiated March 14, 2023

Recall

Recall Number
Z-1429-2023
Event Number
91925
Firm
Hamilton Medical AG Via Crusch 8 Bonaduz Switzerland
FEI Number
3001421318
Product Code
CBK
Status
Open, Classified
Root Cause
Software design
Initiated
March 14, 2023
Posted
April 25, 2023

Description

HAMILTON-C6, REF: 160021

Reason

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

Action

On 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to [email protected] 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Technical Support at 1-800-426-6331, option #2, or email [email protected]

Distribution

US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

Quantity

103