Norco Folding Shower Bench; Features a nonskid surface and handholds for added security. holes allow excess water to drain away easily. Sturdy legs have non-skid rubber feet. Supports up to 200 lbs. North Coast Medical Inc. 8100 Camino Arroyo Gilroy, CA 95020 Comfort and safety in shower
Recall
- Recall Number
- Z-1429-2012
- Event Number
- 61455
- Firm
- North Coast Medical Inc 8100 Camino Arroyo
- FEI Number
- 1000124207
- Product Code
- IKX
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- March 20, 2012
- Posted
- April 16, 2012
- Terminated
- July 20, 2012
- Address
- Gilroy, CA, 95020-7304
Description
Norco Folding Shower Bench; Features a nonskid surface and handholds for added security. holes allow excess water to drain away easily. Sturdy legs have non-skid rubber feet. Supports up to 200 lbs. North Coast Medical Inc. 8100 Camino Arroyo Gilroy, CA 95020 Comfort and safety in shower
Due to a consumer complaint, North Coast Medical determined that due to a design flaw, there was not enough plastic support material to retain bracket bolts when subjected to a force greater than 150 lbs.
North Coast Medical, Inc. sent a Recall Notification Letter dated March 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and discard the affected product. Complete the enclosed Form and return the form even if they no longer have the bath bench. Return the form by fax to 408-775-1430. Customers should contact the Quality Department if they have any questions, concerns or need any assistance following the instructions. For questions regarding this recall call 800-821-9319, ext 121.
Nationwide Distribution
370 benches