Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725), C-PTISYJ-100-HC-G-NA-FLEX7.5 (G57720), C-PTISYJ-100-HC-G-NA-FLEX8.5 (G57721), C-PTISY-100-HC-G-NA-FLEX8.5 (G57717), C-PTISY-100-HC-G-NA-FLEX7.5 (G57716)
Recall
- Recall Number
- Z-1428-2023
- Event Number
- 92083
- Firm
- Cook Incorporated
- FEI Number
- 1820334
- Product Code
- JOH
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 7, 2023
- Posted
- May 10, 2023
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120
Description
Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725), C-PTISYJ-100-HC-G-NA-FLEX7.5 (G57720), C-PTISYJ-100-HC-G-NA-FLEX8.5 (G57721), C-PTISY-100-HC-G-NA-FLEX8.5 (G57717), C-PTISY-100-HC-G-NA-FLEX7.5 (G57716)
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
An URGENT: MEDICAL DEVICE RECALL dated 4/7/23 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets and Trays from the Cook Medical lots in Attachment A. If so, quarantine any affected sets/trays that remain unused. Immediately cease all further distribution and use of the affected sets/trays. 2. Return the complete sets/trays to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected sets/trays on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected sets/trays have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market and replacing potentially defective tracheostomy tubes with new tracheostomy tubes as they become available from Medtronic. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your return of the Acknowledgement and Receipt Form.
Worldwide distribution - US Nationwide.
25,161 units