FDA Recall Open, Classified

HeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.

Recall: Z-1428-2020 · Initiated January 23, 2020

Recall

Recall Number
Z-1428-2020
Event Number
84829
Firm
Heartware, Inc.
FEI Number
3007042319
Product Code
DSQ
Status
Open, Classified
Root Cause
Device Design
Initiated
January 23, 2020
Posted
March 9, 2020
Address
14400 Nw 60th Ave, Miami Lakes, FL, 33014-2807

Description

HeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.

Reason

Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.

Action

Heartware notified customers on about 01/23/2020 via "Urgent Field Safety Notice" letter. Customers were notified of the potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port. When this occurs, communication circuits between the battery and the Controller may be damaged. If circuit damage occurs, it will be necessary to exchange the Controller to restore communication between the batteries and Controller. Instructions included to advise patients to carefully follow the guidance provided in the patient manual related to connecting a power source. Pay attention when connecting and disconnecting power supply cables to ensure users verify the correct cable is being connected to their Controller. Patients were instructed to report any persistent, unexpected audible tones to their Clinician for additional instructions. Additional instructions included to ensure all personnel within your organization or any customers if further distributed are notified of this recall. If you have any questions, please contact the Cardiac and Vascular Group Customer Focused Quality at 763-526-2513 or by email to: [email protected].

Distribution

No affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Taiwan, Province of China, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Quantity

5,489 devices total