Halyard Closed Suction System for Adults, Turbo- cleaningClosedSuctionSystemforAdults,DoubleSwivelElbow (DSE). Used to aspirate liquids or semisolids from a patient's upper airway. a. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54 CM (GREEN); b. HALYARD* TRACH CARE-72 DSE ENDO 14FR; c. HALYARD* TRACH CARE-72 WET PAK* 14FR; d. HALYARD* TRACH CARE-72 WET PAK* T-PIECE SWVL/FLEX ENDO 14FR; e. HALYARD* TURBO-CLEANING 14FR DSE 12IN/30.5 CM (GREEN); f. HALYARD* TRACH CARE-72 DSE TRACH 14FR; g. HALYARD* TRACH CARE-72 WET PAK* DSE TRACH 14FR; h. HALYARD* TRACH CARE-72 DSE ENDO 10FR; i. HALYARD* TURBO-CLEANING 10FR DSE 12IN/30.5CM (BLACK); j. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 10FR; k. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 12FR; l. HALYARD* TURBO-CLEANING 12FR DSE 12IN/30.5CM (WHITE); m. HALYARD* 72 DSE MDI BUILT IN ENDO 12FR; n. HALYARD* 72 DSE TRACH CARE 12FR; o. HALYARD* 14FR DSE 22.2IN/56CM (GREEN); p. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 14FR; q. HALYARD* TURBO-CLEANING 14FR DSE 12IN/30.5CM (GREEN); r. HALYARD* TRACH CARE-72 DSE MDI BUILT IN TRACH 14FR; s. HALYARD* TC-72 WET PAK* DSE MDI ENDO 14FR; t. HALYARD* TC-72 WET PAK DSE MDI ENDO 14FR; u. HALYARD* TC-72 WET PAK DSE 14FR; v. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54CM (GREEN); w. HALYARD* TURBO-CLEANING 12FR DSE 21.3IN/54 CM (WHITE); x. HALYARD* TRACH CARE-72 DSE ENDO 12FR; y. HALYARD* TURBO-CLEANING 12FR DSE 12IN/30.5CM (WHITE); z. HALYARD* TRACH CARE-72 DSE TRACH 12FR; aa. HALYARD* TRACH CARE-72 WET PAK* DSE TRACH 12FR; ab. HALYARD* TURBO-CLEANING 16FR DSE 21.3IN/54CM (ORANGE); ac. HALYARD* TRACH CARE-72 DSE 16FR 21.3IN/54CM; ad. HALYARD* TRACH CARE-72 DSE TRACH 16FR; ae. HALYARD* TRACH CARE-72 DSE ENDO 16FR; af. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 16FR; ag. HALYARD* TRACH CARE-72 DSE MDI BUILT IN TRACH 16FR; ah. HALYARD* TRACH CARE-72 DSE MDI ENDO 16FR; ai. HALYARD* TRACH CARE-72 DSE 16FR; aj. HALYARD* TRACH CARE-72 WET PAK* DSE ENDO 12FR; ak. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54CM (GREEN); al. HALYARD* TURBO-CLEANING 14FR DSE Endo;

Recall: Z-1428-2018 · Initiated February 7, 2018

Recall

Recall Number
Z-1428-2018
Event Number
79531
Firm
Halyard Health, Inc
FEI Number
3011270181
Product Code
BSY
Status
Terminated
Root Cause
Labeling design
Initiated
February 7, 2018
Terminated
January 18, 2022
Address
5405 Windward Pkwy, Alpharetta, GA, 30004-3894

Description

Halyard Closed Suction System for Adults, Turbo- cleaningClosedSuctionSystemforAdults,DoubleSwivelElbow (DSE). Used to aspirate liquids or semisolids from a patient's upper airway. a. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54 CM (GREEN); b. HALYARD* TRACH CARE-72 DSE ENDO 14FR; c. HALYARD* TRACH CARE-72 WET PAK* 14FR; d. HALYARD* TRACH CARE-72 WET PAK* T-PIECE SWVL/FLEX ENDO 14FR; e. HALYARD* TURBO-CLEANING 14FR DSE 12IN/30.5 CM (GREEN); f. HALYARD* TRACH CARE-72 DSE TRACH 14FR; g. HALYARD* TRACH CARE-72 WET PAK* DSE TRACH 14FR; h. HALYARD* TRACH CARE-72 DSE ENDO 10FR; i. HALYARD* TURBO-CLEANING 10FR DSE 12IN/30.5CM (BLACK); j. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 10FR; k. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 12FR; l. HALYARD* TURBO-CLEANING 12FR DSE 12IN/30.5CM (WHITE); m. HALYARD* 72 DSE MDI BUILT IN ENDO 12FR; n. HALYARD* 72 DSE TRACH CARE 12FR; o. HALYARD* 14FR DSE 22.2IN/56CM (GREEN); p. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 14FR; q. HALYARD* TURBO-CLEANING 14FR DSE 12IN/30.5CM (GREEN); r. HALYARD* TRACH CARE-72 DSE MDI BUILT IN TRACH 14FR; s. HALYARD* TC-72 WET PAK* DSE MDI ENDO 14FR; t. HALYARD* TC-72 WET PAK DSE MDI ENDO 14FR; u. HALYARD* TC-72 WET PAK DSE 14FR; v. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54CM (GREEN); w. HALYARD* TURBO-CLEANING 12FR DSE 21.3IN/54 CM (WHITE); x. HALYARD* TRACH CARE-72 DSE ENDO 12FR; y. HALYARD* TURBO-CLEANING 12FR DSE 12IN/30.5CM (WHITE); z. HALYARD* TRACH CARE-72 DSE TRACH 12FR; aa. HALYARD* TRACH CARE-72 WET PAK* DSE TRACH 12FR; ab. HALYARD* TURBO-CLEANING 16FR DSE 21.3IN/54CM (ORANGE); ac. HALYARD* TRACH CARE-72 DSE 16FR 21.3IN/54CM; ad. HALYARD* TRACH CARE-72 DSE TRACH 16FR; ae. HALYARD* TRACH CARE-72 DSE ENDO 16FR; af. HALYARD* TRACH CARE-72 DSE MDI BUILT IN ENDO 16FR; ag. HALYARD* TRACH CARE-72 DSE MDI BUILT IN TRACH 16FR; ah. HALYARD* TRACH CARE-72 DSE MDI ENDO 16FR; ai. HALYARD* TRACH CARE-72 DSE 16FR; aj. HALYARD* TRACH CARE-72 WET PAK* DSE ENDO 12FR; ak. HALYARD* TURBO-CLEANING 14FR DSE 21.3IN/54CM (GREEN); al. HALYARD* TURBO-CLEANING 14FR DSE Endo;

Reason

Certain Flex Connectors supplied with Halyard Closed Suction Kits with Flex Connector may become loose or disconnect before use or during use.

Action

Customers were notified via mail and/or email on about 02/07/2018. Instructions for distributors included to notify all affected customers who may have received potentially impacted products and to complete and return the Distributor Acknowledgement Form. Instructions for customers included to ensure adequate connection of the Flex Connector to Swivel Connector before use of the product by following. This accomplished by ensuring a tight fit is achieved on the Swivel with the Flex Connector advanced approximately three-quarters over the Swivel, so a gap remains visible. If and adequate connection is observed, the Flex Connector may be replaced with a new, sterile Flex Connector and the system connection should be re-evaluated. Lastly, if use of a Flex Connector with the ventilator circuit is optional, the closed suction Swivel Connector can be connected directly to the ventilator circuit. During use, if a disconnection between the Flex Connector and Swivel Connector is observed during use, replace the Flex Connector with a new, sterile Flex Connector (e.g. HYH Product # 1115) and ensure adequate connection as described in this notice. If an inadequate connection occurs, Halyard Health instructs customers to safe the Flex Connector and report it via the Product Complaint Process (PIQ) and the local Halyard Health Sales Representative will assist. Additionally customers were instructed to complete and return the Acknowledgement Form and distribute the recall notice to all clinicians in the facility that may be impacted.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, ARGENTINA, AUSTRALIA, BAHAMAS, BELGIUM, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DOMINICAN REP., ECUADOR, EL SALVADOR, GUATEMALA, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, LEBANON, MALAYSIA, MEXICO, NEW ZEALAND, PERU, PHILIPPINES, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, THAILAND, UNITED KINGDOM, and UNITED ARAB EMIRATES.

Quantity

32,189 cases (20 units per case - 643,780 units) total