FDA Recall Open, Classified

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A33178/ A33178 (5)Torrent Suite Dx Software 5.0- A29166/ A29166

Recall: Z-1426-2024 · Initiated February 7, 2024

Recall

Recall Number
Z-1426-2024
Event Number
94122
Firm
Life Technologies Corporation
FEI Number
3003335080
Product Code
PFF
Status
Open, Classified
Root Cause
Software design
Initiated
February 7, 2024
Posted
March 29, 2024
Address
7335 Executive Way, Frederick, MD, 21704-8354

Description

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A33178/ A33178 (5)Torrent Suite Dx Software 5.0- A29166/ A29166

Reason

Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings, configurations, software, or data on the instrument

Action

Thermo Fisher issued a Medical Device Advisory Notice on Feb 7 2024 to US Customers. Letter includes instructions for mitigation of the vulnerabilities,health risk and action to take: Therefore, Thermo Fisher Scientific advises customers using impacted versions of Torrent Suite Dx to disconnect the Ion PGM Dx sequencer, Ion OneTouch instrument and the Ion PGM" Torrent Server from the customer's network to mitigate the vulnerabilities. To access the data results generated on the instrument, Thermo Fisher Scientific recommends connecting the Ion OneTouch instrument and the Ion PGM Dx Sequencer directly into the Ion PGM Torrent Server. Customers can access the Ion PGM Torrent Server by connecting their laptops (with WiFi disabled) directly into the server. Please have your IT department contact your local Thermo Fisher Scientific field support team for mitigation guidance. Thermo Fisher Scientific advises the user facility to have proper network firewalls and controlled access to laboratory equipment and laboratory systems. We will keep you informed if new or additional information becomes available. Contacting your support representative

Distribution

US Nationwide distribution in the states of CA, IN, NC OR, TN.

Quantity

28 units