FDA Recall Terminated

User Manual addendum, TORS for da Vinci Surgical System IS 1200, IS 2000 and IS 3000. Product Usage: The user manual for Transoral Surgery (TORS) provided information to update and replace the Indications for Use and Representative uses section and add the section Warnings for Transoral Otolaryngology Surgery.

Recall: Z-1424-2013 · Initiated January 24, 2013

Recall

Recall Number
Z-1424-2013
Event Number
65094
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Terminated
Root Cause
Labeling False and Misleading
Initiated
January 24, 2013
Posted
May 29, 2013
Terminated
May 30, 2013
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304

Description

User Manual addendum, TORS for da Vinci Surgical System IS 1200, IS 2000 and IS 3000. Product Usage: The user manual for Transoral Surgery (TORS) provided information to update and replace the Indications for Use and Representative uses section and add the section Warnings for Transoral Otolaryngology Surgery.

Reason

Change to User Manual to remove one indication for use and amend warnings to emphasize restrictions for use.

Action

Intuitive sent an Important Medical Device Correction letters dated January 24, 2013 via Federal Express to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to locate and discard the existing TORS addendum, place the new TORS addendum included with this letter with your System User Manual, and complete and return the attached Acknowledgement Form to Intuitive Surgical. For questions call Intuitive Surgical Customer Service at 1-800-876-1310.

Distribution

US Distribution Only

Quantity

1,374