Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD
Recall
- Recall Number
- Z-1423-2022
- Event Number
- 90471
- Firm
- Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
- FEI Number
- 1000361607
- Product Code
- DDG
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- June 9, 2022
- Terminated
- February 8, 2024
Description
Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
On June 9, 2022, the firm notified customers of the recall via email. The issue only affects Siemens Atellica CH analyzer users. No action is required if the assay is being run on other platforms. The issue will be resolved in Atellica Solution Software (SW) version 1.25.3. Until all Atellica CH analyzers in the laboratory are updated to SW 1.25.3, the assay can be run by Open Channel. Please contact [email protected] for an Atellica CH Open Channel application sheet. If the assay is not run by Open Channel prior to installation of SW version 1.25.3, the assay must be re-calibrated once 1.25.3 is installed. Customers were instructed to review results generated with the affected batches in line with the clinical profile of the patient.
US Nationwide distribution in the state of IL.
12 kits (US)