FDA Recall Terminated

Cobe Optima XP Hollow Fiber Membrane Oxygenator/Reservoir, Sterile, (Standalone Systems) Catalog Numbers 050255500, 050316400, 050412400, 050500000. COBE Cardiovascular, Inc..

Recall: Z-1422-06 · Initiated June 26, 2006

Recall

Recall Number
Z-1422-06
Event Number
35839
Firm
Cobe Cardiovascular, Inc
FEI Number
1718850
Product Code
DTZ
Status
Terminated
Root Cause
Other
Initiated
June 26, 2006
Posted
August 24, 2006
Terminated
August 8, 2011
Address
14401 W 65th Way, Arvada, CO, 80004-3524

Description

Cobe Optima XP Hollow Fiber Membrane Oxygenator/Reservoir, Sterile, (Standalone Systems) Catalog Numbers 050255500, 050316400, 050412400, 050500000. COBE Cardiovascular, Inc..

Reason

Cooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a metalic material, thus decreasing the surface area needed for proper heat transfer.

Action

Consignees were notified by letter on 06/26/2006.

Distribution

Worldwide distribution --- USA including VA facilities in CO, GA, MO and countries of Australia, Canada, Italy, and Japan

Quantity

2204 units (Domestic)