FDA Recall Open, Classified

LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

Recall: Z-1421-2022 · Initiated June 13, 2022

Recall

Recall Number
Z-1421-2022
Event Number
90469
Firm
CooperSurgical, Inc.
FEI Number
1216677
Product Code
MQL
Status
Open, Classified
Root Cause
Error in labeling
Initiated
June 13, 2022
Address
95 Corporate Dr, Trumbull, CT, 06611-1350

Description

LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055

Reason

There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.

Action

On June 13, 2022, the firm distributed Urgent Media Safety Notice letters to affected customers. Customers were informed that setting out a thawing blastocyte for 5 minutes may be detrimental to the survival of the blastocyte. Customers were asked to complete and return an acknowledgement form. Once the firm has received the completed form, they will provide the corrected IFU.

Distribution

Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI, International: Belgium, Canada, Croatia, Netherlands, Portugal, Spain

Quantity

403 units