FDA Recall Terminated

GE Healthcare Centricity Cardiology AI1000 Workstation version 4.2.6 and version 4.2.7.05

Recall: Z-1421-05 · Initiated July 18, 2005

Recall

Recall Number
Z-1421-05
Event Number
32707
Firm
GE Medical Systems Information Technologies
FEI Number
1419826
Product Code
LLZ
Status
Terminated
Root Cause
Other
Initiated
July 18, 2005
Posted
August 30, 2005
Terminated
July 22, 2006
Address
800 E Business Center Dr, Mount Prospect, IL, 60056-2178

Description

GE Healthcare Centricity Cardiology AI1000 Workstation version 4.2.6 and version 4.2.7.05

Reason

The potential exists for the AI1000 workstation to mix images from one study with images from another study. Mixed studies could be from the same patient or from different patients. Mixed patient images could compromise patient condition or diagnosis and treatment.

Action

An Urgent Medical Device Corrrection letter, DATED?, was sent to customers listing procedures to take to minimize the possibility of mixed patient images until a permanent software upgrade is installed. Installation of new software can be done by the customer or by a GE Service Representative. A Customer Reply Form accompanies the letter for customers to indicate the location, serial no., system ID, software version of their AI100 Cardiac Review Workstation; and to indicate their upgrade option, either self or GE Service.

Distribution

Outside the US to Austria, Australia, Canada, France, Germany, Italy, Japan, Korea, Malaysia, Norway, Spain, Switzerland, Taiwan, Thailand and United Kingdom. Within the US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA and WI

Quantity

1,287