AIA-360 Automated Immunoassay Analyzer, Product Code 019945
Recall
- Recall Number
- Z-1420-2020
- Event Number
- 84787
- Firm
- Tosoh Bioscience Inc
- FEI Number
- 3005529799
- Product Code
- KHO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- January 10, 2020
- Posted
- February 11, 2020
- Terminated
- September 10, 2020
- Address
- 3600 Gantz Rd, Grove City, OH, 43123-1895
Description
AIA-360 Automated Immunoassay Analyzer, Product Code 019945
A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.
The firm distributed Urgent Medical Device Correction letters to customers on January 10, 2020. Customers were instructed to retrain from touching the display panel while the analyzer is running. If the analyzer stops and/or the display freezes, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. If your analyzer's serial number is listed among the recalled units, you will be contacted by a Tosoh Field Service Engineer to schedule a visit to replace the display screen which will correct the potential issue. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette OConnell at (800) 248-6764 or by email at [email protected]. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Distributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay