FDA Recall
Open, Classified
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
Recall: Z-1420-2018
·
Initiated March 7, 2018
Recall
- Recall Number
- Z-1420-2018
- Event Number
- 79627
- FEI Number
- 3008926093
- Product Code
- FSA
- Status
- Open, Classified
- Root Cause
- Component design/selection
- Initiated
- March 7, 2018
- Address
- Autochair Wood St North Meadow Lane Industrial Estate Alfreton United Kingdom
Description
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
Reason
Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
Action
Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.
Distribution
PA, and worldwide