Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Recall: Z-1420-2018 · Initiated March 7, 2018

Recall

Recall Number
Z-1420-2018
Event Number
79627
FEI Number
3008926093
Product Code
FSA
Status
Open, Classified
Root Cause
Component design/selection
Initiated
March 7, 2018
Address
Autochair Wood St North Meadow Lane Industrial Estate Alfreton United Kingdom

Description

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Reason

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Action

Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.

Distribution

PA, and worldwide