FDA Recall Terminated

Stryker Navigation System - SpineMap(TM) 3D - Sales Demo SOFTWARE, Part number 81-80045. It provides precise stereostatic determination of surgical targets using stereostatic methodology. The three principle features include computer calculation of stereostatic coordinates from the diagnostic images, measurement of the stereostatic coordinates within the surgical field, high-resolution computer display or diagnostic images with stereostatic coordinates indicated. The system is comprised of software and hardware components.

Recall: Z-1418-2011 · Initiated November 10, 2010

Recall

Recall Number
Z-1418-2011
Event Number
57487
Firm
Stryker Instruments Division of Stryker Corporation
FEI Number
1811755
Product Code
HAW
Status
Terminated
Root Cause
Software design
Initiated
November 10, 2010
Posted
February 25, 2011
Terminated
June 1, 2011
Address
4100 East Milham Ave., Kalamazoo, MI, 49001

Description

Stryker Navigation System - SpineMap(TM) 3D - Sales Demo SOFTWARE, Part number 81-80045. It provides precise stereostatic determination of surgical targets using stereostatic methodology. The three principle features include computer calculation of stereostatic coordinates from the diagnostic images, measurement of the stereostatic coordinates within the surgical field, high-resolution computer display or diagnostic images with stereostatic coordinates indicated. The system is comprised of software and hardware components.

Reason

When users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate correlation results or incorrectly shown orientation of every second image series during navigation.

Action

Urgent: Medical Device Recall Notification letters were sent on 12/3/2010, and software upgrades were to be performed in the field by the end of 2010. Specific Customer Instructions: 1) Locate the units listed in the notification. 2) Distribute this notification to all departments affected. 3) Return the enclosed business reply form to confirm receipt of this notification and identify how many affected units are currently in your inventory. 4) Once the new software is available, Stryker will dispatch a service tech in order to update the software. Note: Software upgrades for all applications will be available by the end of the year. The upgrades will be scheduled as we receive your business reply form. Note: If you currently have a Time License or Pay Per Use card, a new one will be issued to you at the time of the update. 5) If you have loaned or sold any of the units listed in this notification, please, forward a copy of this notice to the new users and advise Stryker of the new users name and their new location. 6) If you have disposed of any of the affected units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial numbers. Questions regarding this recall should be directed to Stryker Field Action Team at 800-800-4236, ext 3808, 3584, or 4354 between the hours of 8 a.m. - 5 p.m. Monday - Friday.

Distribution

Worldwide Distribution -- USA, Canada, Mexico, Australia, Argentina, Brazil, United Kingdom, Japan, Poland, S. Korea, Hong Kong, Taiwan, India, and New Zealand.

Quantity

1,155 total