FDA Recall
Terminated
MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product Usage: intended for the collection and transport of clinical specimens containing viruses, from the collection site to the testing laboratory. I
Recall: Z-1417-2021
·
Initiated February 16, 2021
Recall
- Recall Number
- Z-1417-2021
- Event Number
- 87575
- Firm
- Vectornate USA Inc
- FEI Number
- 3015989005
- Product Code
- JSG
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- February 16, 2021
- Terminated
- December 8, 2022
- Address
- 10 Industrial Ave, Ste 4, Mahwah, NJ, 07430-2284
Description
MedSchenker Smart Transport Medium (STM) System: CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL - MULTIPURPOSE CULTURE MEDIUM - Product Usage: intended for the collection and transport of clinical specimens containing viruses, from the collection site to the testing laboratory. I
Reason
Discoloration and turbidity were present in the specified lots of STMs because of Longer Monsoon season at the South Korean Manufacturing Facility.
Action
The firm did not give out a public statement of a recall or recall strategy as it only affected one client and was specific to two identified lots. the firm did notify the customer of the issue via phone and email.
Distribution
US Nationwide distribution in the state of NJ.
Quantity
130,000 devices