Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) L & D PACK-LF, Model Number: DYNJ0456816O; 6) L & D PACK-LF, Model Number: DYNJ0456816P; 7) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 8) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 10) DELIVERY PACK-LF, Model Number: DYNJ20546F; 11) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 13) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 14) OB PACK-LF, Model Number: DYNJ26121I; 15) OB DELIVERY PACK, Model Number: DYNJ31024J; 16) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 17) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 18) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 19) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 20) VAG DELIVERY PACK, Model Number: DYNJ38481G; 21) VAG DELIVERY PACK, Model Number: DYNJ38481I; 22) VAG DELIVERY PACK, Model Number: DYNJ38481K; 23) OB PACK 2, Model Number: DYNJ39000D; 24) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 25) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 26) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 27) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 28) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 29) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 31) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 32) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 33) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 34) OB-VAG PACK 9128190, Model Number: DYNJ62341; 35) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 36) L & D PACK, Model Number: DYNJ63216A; 37) L&D PACK, Model Number: DYNJ63757A; 38) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 39) MINOR VAG PACK, Model Number: DYNJ68663; 40) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 41) GYN VAGINAL PACK, Model Number: DYNJ68681; 42) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 43) VAGINAL PACK, Model Number: DYNJ69774A; 44) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 45) DELIVERY PACK, Model Number: DYNJ87469A; 46) VAGINAL DELIVERY, Model Number: DYNJ901387A; 47) FLOYD DELIVERY, Model Number: DYNJ906937D; 48) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 49) OB PACK, Model Number: DYNJT5186; 50) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 51) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 52) LABOR AND DELIVERY KIT, Model Number: MNS8950A
Recall
- Recall Number
- Z-1414-2026
- Event Number
- 98329
- FEI Number
- 1417592
- Product Code
- MLS
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- January 7, 2026
- Posted
- February 17, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Medline Convenience Kits: 1) OB KIT, Model Number: DYKM2326; 2) KIT LABOR DELIVERY TUBAL LIGAT, Model Number: DYKMBNDL165; 3) KIT LABOR & DELIVERY D&C/D&E, Model Number: DYKMBNDL31; 4) KIT LABOR & DELIVERY CERCLAGE, Model Number: DYKMBNDL38; 5) L & D PACK-LF, Model Number: DYNJ0456816O; 6) L & D PACK-LF, Model Number: DYNJ0456816P; 7) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ04692S; 8) VAGINAL DELIVERY PACK, Model Number: DYNJ16621B; 9) VAGINAL DELIVERY PACK, Model Number: DYNJ16621C; 10) DELIVERY PACK-LF, Model Number: DYNJ20546F; 11) VAGINAL DELIVERY PK, Model Number: DYNJ20979I; 12) VAGINAL DELIVERY PK, Model Number: DYNJ20979J; 13) VAGINAL DELIVERY PACK, Model Number: DYNJ24108D; 14) OB PACK-LF, Model Number: DYNJ26121I; 15) OB DELIVERY PACK, Model Number: DYNJ31024J; 16) OB DELIVERY PACK-LF, Model Number: DYNJ32139C; 17) DB VAG HYSTERECTOMY-LF, Model Number: DYNJ33590F; 18) MAJOR VAGINAL PACK, Model Number: DYNJ35488F; 19) ENS. ACCOUCHEMENT VAGINAL-LF, Model Number: DYNJ37331C; 20) VAG DELIVERY PACK, Model Number: DYNJ38481G; 21) VAG DELIVERY PACK, Model Number: DYNJ38481I; 22) VAG DELIVERY PACK, Model Number: DYNJ38481K; 23) OB PACK 2, Model Number: DYNJ39000D; 24) GYNECOLOGY VAGINAL KIT, Model Number: DYNJ40203A; 25) LABOR DELIVERY POUCH SCREEN, Model Number: DYNJ40918A; 26) L & D PACK WITHOUT INSTRUMENTS, Model Number: DYNJ44720B; 27) VAGINAL DELIVERY PACK, Model Number: DYNJ51377G; 28) FH VAG DELIVERY PK-LF, Model Number: DYNJ51794B; 29) VAGINAL DELIVERY PACK, Model Number: DYNJ55363A; 30) VAGINAL DELIVERY PACK, Model Number: DYNJ55363B; 31) MINOR VAGINAL PACK, Model Number: DYNJ61269B; 32) OB VAG DELIVERY PACK, Model Number: DYNJ61574; 33) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ61650; 34) OB-VAG PACK 9128190, Model Number: DYNJ62341; 35) HP VAGINAL DELIVERY PACK-LF, Model Number: DYNJ62402A; 36) L & D PACK, Model Number: DYNJ63216A; 37) L&D PACK, Model Number: DYNJ63757A; 38) GYN PACK DYNJ55833A, Model Number: DYNJ65145; 39) MINOR VAG PACK, Model Number: DYNJ68663; 40) UTHET TYLER OB PACK, Model Number: DYNJ68673A; 41) GYN VAGINAL PACK, Model Number: DYNJ68681; 42) VAGINAL DELIVERY PACK, Model Number: DYNJ69544A; 43) VAGINAL PACK, Model Number: DYNJ69774A; 44) LABOR & DELIVERY PACK-LF, Model Number: DYNJ84393; 45) DELIVERY PACK, Model Number: DYNJ87469A; 46) VAGINAL DELIVERY, Model Number: DYNJ901387A; 47) FLOYD DELIVERY, Model Number: DYNJ906937D; 48) OB VAGINAL MEDSTAR PACK-LF, Model Number: DYNJT3445; 49) OB PACK, Model Number: DYNJT5186; 50) VAGINAL DELIVERY KIT, Model Number: MNS13550A; 51) VAGINAL DELIVERY TRAY, Model Number: MNS7460A; 52) LABOR AND DELIVERY KIT, Model Number: MNS8950A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-003 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Medline Kits containing Proxima Sterile Surgical Gowns and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form,
Worldwide distribution - US Nationwide.
27320 units