FDA Recall Open, Classified

Hill-Rom PRO+ MRS Surface For Versacare mattress, REF P3255A02.

Recall: Z-1414-2022 · Initiated June 24, 2022

Recall

Recall Number
Z-1414-2022
Event Number
90511
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
IOQ
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
June 24, 2022
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

Hill-Rom PRO+ MRS Surface For Versacare mattress, REF P3255A02.

Reason

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Action

The recalling firm issued a Medical Device Correction Letter dated 6/24/2022 via UPS on 6/24/2022. The letter described the problem, the potential risk, and the actions to be taken by the user. The actions were to inspect their product and check for any indication of error codes present. Identification of error codes were described depending on the configuration of the consignee's product. If the consignee detects an error code, they are to contact the Hillrom Technical Support to have the PRO+ Surface corrected. A list of affected serial numbers was attached to the letter as Appendix 1. Distributors were instructed to share the notification with their end users. A response form was attached to be completed and returned within 2 weeks to confirm they have reviewed and understand the attached letter and that distributors have shared the notification with their end users and will complete the field action

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. There was government distribution but no military distribution. The country of Canada.

Quantity

202 mattresses