FDA Recall Terminated

Medtronic Intrathecal Catheter, Model 8731, (part of infusion systems using Medtronic implantable pumps) The catheter is designed for use in the intrathecal space. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.

Recall: Z-1414-06 · Initiated July 21, 2006

Recall

Recall Number
Z-1414-06
Event Number
35931
Firm
Medtronic Neurological
FEI Number
2182207
Product Code
LKK
Status
Terminated
Root Cause
Other
Initiated
July 21, 2006
Posted
September 9, 2006
Terminated
June 21, 2007
Address
800 53rd Ave NE PO Box 1250, Minneapolis, MN, 55440-1250

Description

Medtronic Intrathecal Catheter, Model 8731, (part of infusion systems using Medtronic implantable pumps) The catheter is designed for use in the intrathecal space. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.

Reason

Tip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.

Action

An Urgent Worldwide Medical Device Recall letter was sent to affected customers 07/20/06 for delivery on 07/21/06. The letter describes the issue, identifies the affected product and list action required by the customer. Affected devices are to be removed from active inventory, quarantined and returned to Medtronic for a warranty credit. A Reply card is asked to be returned. Medtronic Representatives will follow-up by July 28, 2006.

Distribution

Worldwide Distribution-USA and Austria, Belgium, Canada, Cyprus, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, and the United Kingdom

Quantity

Recalls Z-1414-1415-06-USA - 359 devices. Foreign - 575 devices