FDA Recall Terminated

RSP Impaction Fixture

Recall: Z-1413-2017 · Initiated February 9, 2017

Recall

Recall Number
Z-1413-2017
Event Number
76456
FEI Number
1000116912
Product Code
HWA
Status
Terminated
Root Cause
Device Design
Initiated
February 9, 2017
Terminated
October 20, 2017
Address
9800 Metric Blvd, Encore Medical, Lp, Austin, TX, 78758-5445

Description

RSP Impaction Fixture

Reason

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Action

There are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.

Distribution

US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany

Quantity

626 units