FDA Recall Open, Classified

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Recall: Z-1411-2021 · Initiated March 5, 2021

Recall

Recall Number
Z-1411-2021
Event Number
87542
FEI Number
2647580
Product Code
GEI
Status
Open, Classified
Root Cause
Process design
Initiated
March 5, 2021
Address
Covidien 201 Sabanetas Industrial Park Building 911-67, Ponce, PR, 00716

Description

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Reason

Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

Action

On 3/10/2021, Medtronic issued an Urgent Medical Device Recall notices to customers. All consignees were notified by letter delivered via Federal Express and/or certified mail. The letter informs customers of the recall of specific item codes and production lots of Covidien Surgiwand" II Suction and Irrigation Device following reports from customers of foreign particles in the device tubing.

Distribution

Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Quantity

59,828 devices