FDA Recall Terminated

Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS

Recall: Z-1407-06 · Initiated June 22, 2006

Recall

Recall Number
Z-1407-06
Event Number
35737
Firm
Steris Corporation
FEI Number
1043572
Product Code
FLG
Status
Terminated
Root Cause
Other
Initiated
June 22, 2006
Posted
August 25, 2006
Terminated
January 8, 2007
Address
2720 Gunter Park Dr E, Montgomery, AL, 36109-1410

Description

Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS

Reason

Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiring harness, which may result in smoking, sparking, and fire.

Action

The firm sent out a certified letter to all affected consignees alerting them to problems with the device on 6/22/06.

Distribution

Worldwide USA and countries of Japan, Canada, Hong Kong, Korea, Lebanon, Qatar, Poland, Russia, Saudi Arabia, Singapore, Taiwan, Germany, Northern Mariana Islands, and Malaysia.

Quantity

322 units