FDA Recall Terminated

D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers.

Recall: Z-1406-06 · Initiated January 12, 2005

Recall

Recall Number
Z-1406-06
Event Number
35664
Firm
Radiometer America Inc
FEI Number
1523456
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
January 12, 2005
Posted
August 26, 2006
Terminated
September 14, 2011
Address
810 Sharon Dr, Westlake, OH, 44145-1598

Description

D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers.

Reason

QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.

Action

The firm contacted their customers and informed them of this problem and provided a workaround procedure, in a a letter dated 1/12/2005. A ''Field Action Notification'' (FAN) was issued to the firm''s Field Service personnel on 1/21/2005. The letter instructs their personnel to update the NPT7 application software on all units to V3.05 and to hand over a copy of the letter ''Changes related to the Qualicheck2+ and NPT7,'' to the user/customers. In addition, the firm''s personnel were instructed to remove any of the suspect cartridges found on-hand at each customer location; destroy them;and replace them with different lot numberd product.

Distribution

The recalled device component was distributed to a total of 56 customers, nationwide located in the following states: AL, AR, AZ, CA, CO, GA, ID, IL, KA, MA, ME, MI, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT & WA.

Quantity

545