D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers.
Recall
- Recall Number
- Z-1406-06
- Event Number
- 35664
- Firm
- Radiometer America Inc
- FEI Number
- 1523456
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 12, 2005
- Posted
- August 26, 2006
- Terminated
- September 14, 2011
- Address
- 810 Sharon Dr, Westlake, OH, 44145-1598
Description
D7120, BG-OX Cartridge, D7120 REF 946-003, and 946-001, cuvettes for the NPT7 series of blood gas analyzers.
QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.
The firm contacted their customers and informed them of this problem and provided a workaround procedure, in a a letter dated 1/12/2005. A ''Field Action Notification'' (FAN) was issued to the firm''s Field Service personnel on 1/21/2005. The letter instructs their personnel to update the NPT7 application software on all units to V3.05 and to hand over a copy of the letter ''Changes related to the Qualicheck2+ and NPT7,'' to the user/customers. In addition, the firm''s personnel were instructed to remove any of the suspect cartridges found on-hand at each customer location; destroy them;and replace them with different lot numberd product.
The recalled device component was distributed to a total of 56 customers, nationwide located in the following states: AL, AR, AZ, CA, CO, GA, ID, IL, KA, MA, ME, MI, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT & WA.
545