FDA Recall Open, Classified

CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214

Recall: Z-1405-2025 · Initiated February 14, 2025

Recall

Recall Number
Z-1405-2025
Event Number
96238
Firm
Smiths Medical ASD, Inc.
FEI Number
3012307300
Product Code
CCK
Status
Open, Classified
Root Cause
Process control
Initiated
February 14, 2025
Posted
March 20, 2025
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690

Description

CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214

Reason

Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.

Action

On February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.

Distribution

US Nationwide distribution. International distribution pending.

Quantity

175,994 US; 45,223 OUS