FDA Recall Terminated

Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.

Recall: Z-1405-06 · Initiated June 30, 2006

Recall

Recall Number
Z-1405-06
Event Number
34487
Firm
Terumo Cardiovascular Systems Corp
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
June 30, 2006
Posted
August 22, 2006
Terminated
June 19, 2008
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.

Reason

The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.

Action

The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.

Distribution

Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Quantity

Recall Numbers Z-1403-1405-2006 -1,135 systems in the U.S. plus 723 internationally.