Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.
Recall
- Recall Number
- Z-1405-06
- Event Number
- 34487
- Firm
- Terumo Cardiovascular Systems Corp
- FEI Number
- 1828100
- Product Code
- DTQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 30, 2006
- Posted
- August 22, 2006
- Terminated
- June 19, 2008
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103-9586
Description
Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog No. 801046.
The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.
The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.
Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Recall Numbers Z-1403-1405-2006 -1,135 systems in the U.S. plus 723 internationally.