Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA
Recall
- Recall Number
- Z-1404-2010
- Event Number
- 53031
- Firm
- Novasys Medical Inc
- FEI Number
- 3003647794
- Product Code
- NVJ
- Status
- Terminated
- Root Cause
- Labeling design
- Initiated
- July 27, 2009
- Posted
- April 16, 2010
- Terminated
- January 8, 2011
- Address
- 39684 Eureka Dr, Newark, CA, 94560-4805
Description
Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA
Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects. Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation
A Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures. The company also updated physician information brochures and patient/physician DVDs to include the information. This was shipped to physicians in August/September of 2009. Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777. Questions should be addressed to the firm at the same telephone number.
US nationwide only. (TX, FL, CA, LA, KY, MO, GA,OR WA, HI, AZ, OK AL, TN) .
As of 6/12/09 there were 436 Probes at customer accounts.