PTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcare professionals and individuals with diabetes to measure glucose in whole blood
Recall
- Recall Number
- Z-1403-2013
- Event Number
- 65067
- Firm
- Polymer Technology Systems, Inc.
- FEI Number
- 3015525148
- Product Code
- NBW
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- April 4, 2013
- Posted
- May 25, 2013
- Terminated
- April 16, 2014
- Address
- 7736 Zionsville Rd, Indianapolis, IN, 46268-2175
Description
PTS PANELS, Glucose Test Strips for use with CardioChek Brand Analyzers Product Usage: used by healthcare professionals and individuals with diabetes to measure glucose in whole blood
Internal investigation demonstrated under-recovery for glucose when tested against a reference method. This could result in a low to moderate health risk.
Polymer Technology Systems sent a Field Correction Action letters dated April 4, 2013 to distributors and customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use and dispose of remaining inventory of recalled product. Customers were asked to acknowledge and confirm recall directions were follwed by fax to 1-317-870-5608.
Worldwide Distribution - USA Nationwide including Puerto Rico and countries of: UK, CHL, ENG, BEL, ITA, TKY, BEL, LIT, LTV, EGY, SPG, HOK, SWE, MEX, UAE, COR, ISR, MAL, AUS, POR, POL, SLN and STQ.
7048 total, 3499 in USA