FDA Recall
Terminated
MAGNETOM Verio Nuclear magnetic resonance imaging
Recall: Z-1401-2011
·
Initiated January 10, 2011
Recall
- Recall Number
- Z-1401-2011
- Event Number
- 57781
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 10, 2011
- Posted
- February 23, 2011
- Terminated
- September 25, 2012
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
MAGNETOM Verio Nuclear magnetic resonance imaging
Reason
On some systems the gradient cable connections did not meet firm's specifications after installation
Action
Siemens Medical Solutions USA, Inc sent customers a letter on January 10, 2011 notifying them of an Update Instruction and informing them that recalling firm would preventatively check the gradient cable connections of all systems. For questions regarding this recall call 610-219-6300.
Distribution
Nationwide Distribution including AK, AZ, CA, CT, FL, HI, IL, IN, KY, MA, MD, MI, MN, MO, MT, NC, NJ, NM, NV, OH, OK, OR, PA, RI, TN, TX, VA, WI, and WV
Quantity
92