LANTIS XLINK (Impac MultiAccess Sequencer) on Varian accelerators with the MLC3PI interface
Recall: Z-1401-06
·
Initiated July 7, 2006
Recall
- Recall Number
- Z-1401-06
- Event Number
- 35982
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 7, 2006
- Posted
- August 11, 2006
- Terminated
- March 26, 2008
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
LANTIS XLINK (Impac MultiAccess Sequencer) on Varian accelerators with the MLC3PI interface
Reason
Improper treatment may be delivered if the machine configuration file is edited/set for the Varian MLC revision H file format.
Action
The firm has issued a safety advisory letter to all affected consignees, and will distribute an updated software version as soon as it is available.
Distribution
Worldwide Distribution --- USA and countries including Australia, China, Denmark, Germany, New Zealand, Portugal, Russia, Singapore, South Africa, Sweden and Switzerland
Quantity
23 units