FDA Recall Terminated

Quick Connect Adapter Model #CSS-QA-Adapter MaxTorque(R) Maxtorque Quick Connect Lot #s: 1014001 & 1101011, QTY 1 NON STERILE MAT:17-4PH The intended use: Used to stabilize and aid in the fixation of fractures, fusions, and osteotomies.

Recall: Z-1399-2010 · Initiated October 16, 2009

Recall

Recall Number
Z-1399-2010
Event Number
54542
Firm
Orthohelix Surgical Designs Inc
FEI Number
3000718467
Product Code
LXH
Status
Terminated
Root Cause
Other
Initiated
October 16, 2009
Posted
April 15, 2010
Terminated
April 11, 2012
Address
1065 Medina Rd., Suite 500, Medina, OH, 44256

Description

Quick Connect Adapter Model #CSS-QA-Adapter MaxTorque(R) Maxtorque Quick Connect Lot #s: 1014001 & 1101011, QTY 1 NON STERILE MAT:17-4PH The intended use: Used to stabilize and aid in the fixation of fractures, fusions, and osteotomies.

Reason

The adapters are able to come apart when a retaining ring in the proximal end of the adapter can dislodge from the retaining ring groove and the two ends of the adapter fall apart and are unable to be reassembled without special tools and knowledge of the instrument. Also the adapters have a wobble between the two ends. This is caused when the distal and proximal ends of the adapter assembly were

Action

Two "IMPORTANT RECALL NOTICE" letters, one dated October 16, 2009, and the second letter dated November 19, 2009, were sent to the customers. The letters described the product, problem and action to be taken by the customers. The customer are requested to discontinue use of the adapters and return them to the firm, Orthohelix Surgical Designs. The firm will replace the adapters as soon as possible. Should you have any questions about the systems please contact Kristin Wolff at 330-247-1445 or Jim Bragg at 330-247-1444.

Distribution

Nationwide distribution: CA, CO, FL, GA, IL, KY, MD, MI, MO, MS, NC, NJ, NM, NY, OH, OR, PA, SD, TN, TX, VA, WA , and WI.

Quantity

127 pieces