FDA Recall Open, Classified

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Recall: Z-1396-2025 · Initiated February 19, 2025

Recall

Recall Number
Z-1396-2025
Event Number
96340
Firm
Calyxo, Inc.
FEI Number
3014683069
Product Code
FED
Status
Open, Classified
Root Cause
Device Design
Initiated
February 19, 2025
Posted
March 21, 2025
Address
4473 Willow Rd, Pleasanton, CA, 94588-8571

Description

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Reason

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

Action

On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing [email protected], or calling 833-214-3354 and providing your name, title, institution and address.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.

Quantity

11,246