GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System
Recall
- Recall Number
- Z-1395-2023
- Event Number
- 92051
- Firm
- GRAMERCY EXTREMITY ORTHOPEDICS
- FEI Number
- 3012478585
- Product Code
- HRS
- Status
- Open, Classified
- Root Cause
- Mixed-up of materials/components
- Initiated
- April 4, 2023
- Posted
- April 17, 2023
- Address
- 1239 N Glenville Dr, Richardson, TX, 75081-2412
Description
GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System
Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.
On 4/4/23, recall notices were mailed to customers and distributors who were asked to do the following: 1) Quarantine affected kits so they are ready for removal by the firm s sales reps or other contact. 2) If devices were transferred or further distributed, notify customers, or provide a customer list, so the recalling firm can notify these customers directly. 3) Complete and return the response form to [email protected] Customers with questions can contact the firm at 855-436-2278, 9AM 5PM, Monday - Friday.
US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.
24