GE Discovery MR950 MRI system
Recall
- Recall Number
- Z-1395-2015
- Event Number
- 70701
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- IWE
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- October 24, 2014
- Posted
- May 1, 2015
- Terminated
- May 6, 2015
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Discovery MR950 MRI system
The alignment lasers are missing the labels required by radiation safety regulations.
GE Planned action(s) to repair defect or to bring product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 5 units consists of the following: 1. New labels will be mailed to all purchasers, with instructions to apply the label in the appropriate location on the unit, or a GE field engineer will be sent to each user site to complete the correction. 2. All mailings and service calls will be made at no cost to the purchaser. 3. The corrections will be completed by August 1, 2015. If you have any questions, please contact your local Service Representative or call 1 (800) 437-1171.
US Distribution to the state of CA., and Internationally to Italy and Japan.
5 units (2 in US)