SMV FX-80 Nuclear Camera
Recall
- Recall Number
- Z-1394-04
- Event Number
- 29756
- Firm
- General Electric Med Systems
- FEI Number
- 2126677
- Product Code
- IYX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- July 30, 2004
- Posted
- August 24, 2004
- Terminated
- July 24, 2006
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
SMV FX-80 Nuclear Camera
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
Inside the US. Manufactured and distributed from 1993 - 2000.