SMV FX-80 Nuclear Camera

Recall: Z-1394-04 · Initiated July 30, 2004

Recall

Recall Number
Z-1394-04
Event Number
29756
Firm
General Electric Med Systems
FEI Number
2126677
Product Code
IYX
Status
Terminated
Root Cause
Other
Initiated
July 30, 2004
Posted
August 24, 2004
Terminated
July 24, 2006
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

SMV FX-80 Nuclear Camera

Reason

The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.

Action

The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.

Distribution

Inside the US. Manufactured and distributed from 1993 - 2000.

Quantity

41