Solid state x-ray imager (flat panel/digital imager) Solid state x-ray imager (flat panel/digital imager) Medical
Recall: Z-1391-2018
·
Initiated January 12, 2018
Recall
- Recall Number
- Z-1391-2018
- Event Number
- 79628
- Firm
- NeuroLogica Corporation
- FEI Number
- 3004938766
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- January 12, 2018
- Terminated
- August 14, 2018
- Address
- 14 Electronics Ave, Danvers, MA, 01923-1011
Description
Solid state x-ray imager (flat panel/digital imager) Solid state x-ray imager (flat panel/digital imager) Medical
Reason
While transporting the portable machine with the retrofit equipment in a backwards motion, there is potential for foot injury from the underside of the detector bin of the retrofit unit.
Action
A field safety notice was sent stating that the manufacturer, without charge, will remedy the defect. The notice stated that field upgrades would be performed by December 31, 2017. The notice provided instructions to end users and to distributors. For further questions, please call (978) 564-8503.
Distribution
US Distribution . Only one consignee was shipped 16 units of the affected device.
Quantity
16 units