FDA Recall Terminated

Solid state x-ray imager (flat panel/digital imager) Solid state x-ray imager (flat panel/digital imager) Medical

Recall: Z-1391-2018 · Initiated January 12, 2018

Recall

Recall Number
Z-1391-2018
Event Number
79628
Firm
NeuroLogica Corporation
FEI Number
3004938766
Product Code
MQB
Status
Terminated
Root Cause
Device Design
Initiated
January 12, 2018
Terminated
August 14, 2018
Address
14 Electronics Ave, Danvers, MA, 01923-1011

Description

Solid state x-ray imager (flat panel/digital imager) Solid state x-ray imager (flat panel/digital imager) Medical

Reason

While transporting the portable machine with the retrofit equipment in a backwards motion, there is potential for foot injury from the underside of the detector bin of the retrofit unit.

Action

A field safety notice was sent stating that the manufacturer, without charge, will remedy the defect. The notice stated that field upgrades would be performed by December 31, 2017. The notice provided instructions to end users and to distributors. For further questions, please call (978) 564-8503.

Distribution

US Distribution . Only one consignee was shipped 16 units of the affected device.

Quantity

16 units