FDA Recall Open, Classified

Medline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4A; 4) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4B; 5) VEIN PROCEDURE KIT, Model Number: DYNDA2076C; 6) CARDIAC CATH PACK-LF, Model Number: DYNJ0150605AC; 7) CATH LAB PACK-LF, Model Number: DYNJ0648473Y; 8) NBMC CATH PACK-LF, Model Number: DYNJ0803367I; 9) ANGIO PACK-LF, Model Number: DYNJ0954970I; 10) CARDIAC CATH PACK, Model Number: DYNJ19946M; 11) CATH LAB PACK, Model Number: DYNJ30955I; 12) CATH LAB PACK, Model Number: DYNJ31773K; 13) CATH PACK-LF, Model Number: DYNJ33061D; 14) DB CCL OAKWOOD PACK-LF, Model Number: DYNJ33620D; 15) CARDIAC CATH PACK MRH-LF, Model Number: DYNJ35300M; 16) CATH LAB PK, Model Number: DYNJ35643C; 17) CARDIAC CATH PACK-LF, Model Number: DYNJ39066B; 18) CATH PACK-LF, Model Number: DYNJ43101C; 19) OR ANGIO PACK-LF, Model Number: DYNJ43415D; 20) OR ANGIO PACK-LF, Model Number: DYNJ43415F; 21) ANGIOPLASTY PACK, Model Number: DYNJ44700J; 22) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ44850J; 23) RADIOLOGICAL PACK, Model Number: DYNJ44889O; 24) CARDIAC CATH PACK, Model Number: DYNJ47370B; 25) ANGIOGRAPHY PACK, Model Number: DYNJ53129D; 26) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ56772G; 27) DR. MOHINDRA PACK, Model Number: DYNJ57602G; 28) NE ENDO VASCULAR PACK, Model Number: DYNJ57940B; 29) CARDIAC CATH PACK - Q1, Model Number: DYNJ58048C; 30) OR ANGIO, Model Number: DYNJ59931B; 31) LPCH PERRY CATH PACK, Model Number: DYNJ67378C; 32) HOSPITAL REGIONAL DE CONCEPCIO, Model Number: DYNJ67918A; 33) CATH LAB PACK, Model Number: DYNJ68633; 34) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197B; 35) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197C; 36) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197D; 37) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197F; 38) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197G; 39) BASIC CARDIAC CATH PACK, Model Number: DYNJ81593B; 40) UMC EL PASO CATH PACK, Model Number: DYNJ83089; 41) DR. MOHINDRA PACK, Model Number: DYNJ83819A; 42) VZ VAS/ANG PACK, Model Number: DYNJ83981; 43) WCPN CARDIOLOGY PACK, Model Number: DYNJ84539; 44) CATH LAB, Model Number: DYNJ85511; 45) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJT1520K; 46) ANGIOGRAPHY TRAY, Model Number: MNS9130; 47) CVC BUNDLE KIT, Model Number: ECVC6225A; 48) CVC 3L 7F 20CM VANTEX, Model Number: ECVC7265A

Recall: Z-1389-2026 · Initiated January 7, 2026

Recall

Recall Number
Z-1389-2026
Event Number
98329
FEI Number
1417592
Product Code
OES
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 7, 2026
Posted
February 17, 2026
Address
3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753

Description

Medline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4A; 4) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4B; 5) VEIN PROCEDURE KIT, Model Number: DYNDA2076C; 6) CARDIAC CATH PACK-LF, Model Number: DYNJ0150605AC; 7) CATH LAB PACK-LF, Model Number: DYNJ0648473Y; 8) NBMC CATH PACK-LF, Model Number: DYNJ0803367I; 9) ANGIO PACK-LF, Model Number: DYNJ0954970I; 10) CARDIAC CATH PACK, Model Number: DYNJ19946M; 11) CATH LAB PACK, Model Number: DYNJ30955I; 12) CATH LAB PACK, Model Number: DYNJ31773K; 13) CATH PACK-LF, Model Number: DYNJ33061D; 14) DB CCL OAKWOOD PACK-LF, Model Number: DYNJ33620D; 15) CARDIAC CATH PACK MRH-LF, Model Number: DYNJ35300M; 16) CATH LAB PK, Model Number: DYNJ35643C; 17) CARDIAC CATH PACK-LF, Model Number: DYNJ39066B; 18) CATH PACK-LF, Model Number: DYNJ43101C; 19) OR ANGIO PACK-LF, Model Number: DYNJ43415D; 20) OR ANGIO PACK-LF, Model Number: DYNJ43415F; 21) ANGIOPLASTY PACK, Model Number: DYNJ44700J; 22) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ44850J; 23) RADIOLOGICAL PACK, Model Number: DYNJ44889O; 24) CARDIAC CATH PACK, Model Number: DYNJ47370B; 25) ANGIOGRAPHY PACK, Model Number: DYNJ53129D; 26) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ56772G; 27) DR. MOHINDRA PACK, Model Number: DYNJ57602G; 28) NE ENDO VASCULAR PACK, Model Number: DYNJ57940B; 29) CARDIAC CATH PACK - Q1, Model Number: DYNJ58048C; 30) OR ANGIO, Model Number: DYNJ59931B; 31) LPCH PERRY CATH PACK, Model Number: DYNJ67378C; 32) HOSPITAL REGIONAL DE CONCEPCIO, Model Number: DYNJ67918A; 33) CATH LAB PACK, Model Number: DYNJ68633; 34) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197B; 35) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197C; 36) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197D; 37) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197F; 38) STEWARD CATH LAB PACK GSS, Model Number: DYNJ81197G; 39) BASIC CARDIAC CATH PACK, Model Number: DYNJ81593B; 40) UMC EL PASO CATH PACK, Model Number: DYNJ83089; 41) DR. MOHINDRA PACK, Model Number: DYNJ83819A; 42) VZ VAS/ANG PACK, Model Number: DYNJ83981; 43) WCPN CARDIOLOGY PACK, Model Number: DYNJ84539; 44) CATH LAB, Model Number: DYNJ85511; 45) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJT1520K; 46) ANGIOGRAPHY TRAY, Model Number: MNS9130; 47) CVC BUNDLE KIT, Model Number: ECVC6225A; 48) CVC 3L 7F 20CM VANTEX, Model Number: ECVC7265A

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-003 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Medline Kits containing Proxima Sterile Surgical Gowns and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form,

Distribution

Worldwide distribution - US Nationwide.

Quantity

30958 units