Epix Universal Clip Applier, REF: CA500, STERILEEO included within the following kit Reference Numbers: GK1003, AME Kit 54562 GK1004, KIT LAP CHOL DEN HOED 2, GK1005, K2550 - HPATECTOMIE ERASME, GK1011, K-19459, GK1014, K-18031 STANDARD KIT, GK1018, K-20735 STANDARD KIT, GK1033, AME STANDARD KIT 21-023, GK1044, KE39830 COLE, GK1047, AME STANDARD KIT 22-003, GK1063, AME STANDARD KIT 22-022, GK1068, AME STANDARD KIT 22-027, GK1071, AME STANDARD KIT 22-030, GK1083, AME STANDARD KIT 22-045, GK1094, AME STANDARD KIT 23-015, GK1096, AME STANDARD KIT 23-017, GK1101, AME STANDARD KIT 23-033, GK131, LAP CHOLE PROCEDURE PACK, GK153, TOTAL 5mm SOLUTION KIT 11, GK168, 10mm ACCESS/LIGATION/RETRIEVAL PK, GK258, AME STANDARD KIT 44, GK260, STANDARD SET 360 Z, GK307, CHOLE KIT 1, GK308, CHOLE KIT 15, GK320, CHOLE KIT 24, GK358, CHOLE KIT 22, GK360, CHOLE KIT 18, GK363, CHOLE KIT 31, GK364, CHOLE KIT 30, GK365, CHOLE KIT 28, GK368, CHOLE KIT 33, GK401, TOTAL 5mm SOLUTION KIT 41, GK402, TOTAL 5mm SOLUTION KIT 42, GK736, LAP. CHOLECYST. CLIP, GK738, STANDARD SET 112 F, GK739, STANDARD SET 397 F, GK834, K100796, GK835, KIT AME 55003, GK837, KIT K86656, GK857, KIT CHOLECYSTECTOMIE LOBBES, GK858, KIT KE284, GK861, KIT VESICULE 11MM CHA, GK874, LAP CHOLE KIT Z-THREAD, GK877, LAP CHOL KIT 5, GK878, SSMC LAP, CHOLE, KIT NO.1, GK879, SSMC LAP, CHOLE, KIT NO. 2, GK891, LAP-GALLE-SET, GK895, KIT HERES LAP CHOL 5MM, GK905, GALLKIT TORSBY SJUKHUS, GK920, K57569, GK921, K57526, GK924, K65251, GK962, LAP CHOL BOOM, GK965, URO- VARIKOCELEN+LYMPHEKTOMIESET, GK969, KIT LAP CHOL TOORENVLIET, GK980, KE78892 LAP PROSTATECTOMY 2, GK983, KIT KE 111515, GK987, LAP. CHOLE-SET LKH MURTAL, GK988, LAPCHOLE KIT ERASME VERSION 4, GK992, K8056 - KIT COLE, GK993, K116634, GK998, GALLEKIT VOSS, JK001, OLYMPUS KIT 1 (2XCTF03,1XCA500), JK003, OLYMPUS KIT 3 (2XCFF05, 1XCA500), JK004, OLYMPUS KIT 4 (2XCTR05, 1XCA500), K0880, DR. THAMES-LAP CHOLE, K0924, DR. COMPTON'S LAP CHOLE PACK, K2212, DR. MILLER LAP CHOLE KIT, K2440, LAP CHOLE KIT, K2607, BORLAND GASTRIC SLEEVE, K2640, LAP CHOLE KIT, K2656, SOUTH SUNFLOWER LAP CHOLE KIT, K2703, DR. AZAR LAP CHOLE PACK, K2705, DRS.GIROD & INGRAM LAP CHOLE PACK, K2797, LAP CHOLE PROCEDURAL PACK 1, K2903, LAP CHOLE KIT, K2922, KIT CLIP APPLIER, K2924, CLIP KIT A, K2925, CLIP KIT F, K2926, CLIP KIT F, K2927, CLIP KIT B, K2928, LAP CHOLE KIT, K2929, LAP CHOLE KIT, K2933, AM MENAT KIT # 0001, K2934, AM MENAT KIT # 0002

Recall: Z-1389-2024 · Initiated January 30, 2024

Recall

Recall Number
Z-1389-2024
Event Number
94034
Firm
Applied Medical Resources Corp 22872 Avenida Empresa Rancho Santa
FEI Number
2027111
Product Code
MDM
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 30, 2024
Posted
March 27, 2024
Address
Margarita, CA, 92688-2650

Description

Epix Universal Clip Applier, REF: CA500, STERILEEO included within the following kit Reference Numbers: GK1003, AME Kit 54562 GK1004, KIT LAP CHOL DEN HOED 2, GK1005, K2550 - HPATECTOMIE ERASME, GK1011, K-19459, GK1014, K-18031 STANDARD KIT, GK1018, K-20735 STANDARD KIT, GK1033, AME STANDARD KIT 21-023, GK1044, KE39830 COLE, GK1047, AME STANDARD KIT 22-003, GK1063, AME STANDARD KIT 22-022, GK1068, AME STANDARD KIT 22-027, GK1071, AME STANDARD KIT 22-030, GK1083, AME STANDARD KIT 22-045, GK1094, AME STANDARD KIT 23-015, GK1096, AME STANDARD KIT 23-017, GK1101, AME STANDARD KIT 23-033, GK131, LAP CHOLE PROCEDURE PACK, GK153, TOTAL 5mm SOLUTION KIT 11, GK168, 10mm ACCESS/LIGATION/RETRIEVAL PK, GK258, AME STANDARD KIT 44, GK260, STANDARD SET 360 Z, GK307, CHOLE KIT 1, GK308, CHOLE KIT 15, GK320, CHOLE KIT 24, GK358, CHOLE KIT 22, GK360, CHOLE KIT 18, GK363, CHOLE KIT 31, GK364, CHOLE KIT 30, GK365, CHOLE KIT 28, GK368, CHOLE KIT 33, GK401, TOTAL 5mm SOLUTION KIT 41, GK402, TOTAL 5mm SOLUTION KIT 42, GK736, LAP. CHOLECYST. CLIP, GK738, STANDARD SET 112 F, GK739, STANDARD SET 397 F, GK834, K100796, GK835, KIT AME 55003, GK837, KIT K86656, GK857, KIT CHOLECYSTECTOMIE LOBBES, GK858, KIT KE284, GK861, KIT VESICULE 11MM CHA, GK874, LAP CHOLE KIT Z-THREAD, GK877, LAP CHOL KIT 5, GK878, SSMC LAP, CHOLE, KIT NO.1, GK879, SSMC LAP, CHOLE, KIT NO. 2, GK891, LAP-GALLE-SET, GK895, KIT HERES LAP CHOL 5MM, GK905, GALLKIT TORSBY SJUKHUS, GK920, K57569, GK921, K57526, GK924, K65251, GK962, LAP CHOL BOOM, GK965, URO- VARIKOCELEN+LYMPHEKTOMIESET, GK969, KIT LAP CHOL TOORENVLIET, GK980, KE78892 LAP PROSTATECTOMY 2, GK983, KIT KE 111515, GK987, LAP. CHOLE-SET LKH MURTAL, GK988, LAPCHOLE KIT ERASME VERSION 4, GK992, K8056 - KIT COLE, GK993, K116634, GK998, GALLEKIT VOSS, JK001, OLYMPUS KIT 1 (2XCTF03,1XCA500), JK003, OLYMPUS KIT 3 (2XCFF05, 1XCA500), JK004, OLYMPUS KIT 4 (2XCTR05, 1XCA500), K0880, DR. THAMES-LAP CHOLE, K0924, DR. COMPTON'S LAP CHOLE PACK, K2212, DR. MILLER LAP CHOLE KIT, K2440, LAP CHOLE KIT, K2607, BORLAND GASTRIC SLEEVE, K2640, LAP CHOLE KIT, K2656, SOUTH SUNFLOWER LAP CHOLE KIT, K2703, DR. AZAR LAP CHOLE PACK, K2705, DRS.GIROD & INGRAM LAP CHOLE PACK, K2797, LAP CHOLE PROCEDURAL PACK 1, K2903, LAP CHOLE KIT, K2922, KIT CLIP APPLIER, K2924, CLIP KIT A, K2925, CLIP KIT F, K2926, CLIP KIT F, K2927, CLIP KIT B, K2928, LAP CHOLE KIT, K2929, LAP CHOLE KIT, K2933, AM MENAT KIT # 0001, K2934, AM MENAT KIT # 0002

Reason

Their is the potential that Universal Clip applier may not load clip after the trigger is actuated.

Action

On February 16, 2024, Applied Medical issued a "Urgent: Medical Device Recall" notification to affected consignees via UPS. In addition, to informing consignees about the recall, Applied Medical asked consignees to take the following actions: " Check your inventory for recalled product. " Quarantine any product from the impacted lots above (Page 1) or kits listed below (Page 7-10). " Complete the attached Distributor Recall Notification Confirmation Form (Page 4) to acknowledge the recall and indicate if your facility is returning or has already used the lots listed above. o If no product is being returned, please indicate on the Distributor Recall Notification Confirmation Form (Page 4). o Please notify any facilities to which you distributed the affected product with a copy of this notification. Please also complete the Distributor Recall Notification Confirmation Report (Page 5). o For any product returned to your facility, please return the affected product to Applied Medical. " Return the Recall Notification Confirmation Form to Applied Medical by emailing to [email protected] or faxing to (949) 221-3639. " Return affected product and a copy of the Recall Notification Confirmation Form to Applied Medical (Product Return Instructions are on Page 6).

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Venezuela.

Quantity

17,990 units