FDA Recall Open, Classified

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

Recall: Z-1388-2025 · Initiated February 5, 2025

Recall

Recall Number
Z-1388-2025
Event Number
96256
Firm
Beckman Coulter, Inc.
FEI Number
2122870
Product Code
JJE
Status
Open, Classified
Root Cause
Process design
Initiated
February 5, 2025
Posted
March 17, 2025
Address
1000 Lake Hazeltine Dr, Chaska, MN, 55318-1037

Description

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.

Reason

the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 1/30/25 was sent to customers. Affected DxI 9000 Access Immunoassay Analyzers could experience an increase in the following errors: IncubateWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) WashWheel.Axis MachineControlStatus(Encoder reports that axis position is out of tolerance) This can cause the analyzer to enter the red, or Stopped, system status and any in-process tests are cancelled. There could be a delay in reporting patient test results to the physician. If the error occurs: 1. Confirm that the DxI 9000 Access Immunoassay Analyzer is in either the Paused or Stopped system state. 2. Select the System state button in the lower left corner of the user interface. 3. Select Initialize. When initialization is complete, the analyzer returns to the green, or Running, state. If your laboratory has experienced the issue and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. Beckman Coulter has identified the root cause of this issue and is no longer distributing affected motors for the Incubate and Wash Wheel assemblies. A Beckman Coulter service representative will schedule a visit to replace any affected motors on your DxI 9000 Access Immunoassay Analyzer. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; From our website: http://www.beckmancoulter.com

Distribution

US Nationwide. Global Distribution.

Quantity

206 units